Reinsert the cuvette into the instrument.
Press Read to start the reading. The instrument displays the results in mg/L of free chlorine.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Free Chlorine