1Instruction ManualHI 83206Multiparameter BenchPhotometerfor Environmental Testingwww.hannainst.comThis Instrument is inCompliance with the CE Directi
18 1910 mL• Remove the cuvette.• Add 4 drops of HI 93715A-0 First reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4
20 21FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3%
22 23• Reinsert the cuvette into the instrument.• Press READ to start the reading. The instrument displays the resultsin mg/L of free chlorine.INTERFE
24 25Total Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen
26 27Chromium VI HRCHROMIUM VI HIGH RANGESPECIFICATIONSRange 0 to 1000 µg/LResolution 1 µg/LAccuracy ±5 µg/L ±4% of readingTypical EMC ±1 µg/LDeviatio
28 29Chromium VI LRCHROMIUM VI LOW RANGESPECIFICATIONSRange 0 to 300 µg/LResolution 1 µg/LAccuracy ±1 µg/L ±4% of readingTypical EMC ±1 µg/LDeviationL
30 31Color of WaterCOLOR OF WATERSPECIFICATIONSRange 0 to 500 PCU (Platinum Cobalt Units)Resolution 1 PCUAccuracy ±10 PCU ±5% of readingTypical EMC ±
32 33Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviat
34 35Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of readingTypical EMC ±1 μg/LDeviationLight Sou
36 37Cyanuric AcidCYANURIC ACIDSPECIFICATIONSRange 0 to 80 mg/LResolution 1 mg/LAccuracy ±1 mg/L ±15% of readingTypical EMC ±1 mg/LDeviationLight Sour
2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,
38 39MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight S
40 41Molybdenum• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, waitfor 5 minutes and press READ. Whe
42 43Nickel HR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternative
44 45Nickel LR• Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of H
46 47NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungst
48 49Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of readingTypical EMC ±1 mg/LDeviationLight So
50 51SPECIFICATIONSRange 0.00 to 0.35 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungsten l
52 53Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of readingTypical EMC ± 0.1 mg/LDeviationLight Sour
54 55pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pHTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow band int
56 57Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of readingTypical EMC Dev. ±0.1 mg/LL
4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave
58 59Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of readingTypical EMC Dev. ±0.01
60 61PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of readingTypical EMC Dev. ±0.2 mg/LLight Sourc
62 63SilicaSILICASPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Sourc
64 65INTERFERENCESInterference may be caused by:Phosphate above 60 mg/LPhosphate above 75 mg/LSulfide and high concentration of ironEliminate color an
66 67Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker(the sample) and swirl gently to mix.• Press TIMER and the display w
68 69ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tu
70 71The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messag
72 73Description Range MethodAmmonia MR 0.00 to 10.00 mg /L NesslerAmmonia LR 0.00 to 3.00 mg /L NesslerChlorine, Free 0.00 to 2.50 mg /L DPDChlorine,
74 75WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose
76Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@vi
6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con
8 9USING LIQUID AND POWDER REAGENTS• Proper use of the dropper:(a) for reproducible results, tap the dropper on the table for several times and wipe
10 11Pay attention to push the cuvette completely down in the holder.• In order to avoid reagent leaking and to obtain more accurate measure-ments, cl
12 13POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:
14 15Date / TimeThis option is used to set the instrument’s date and time.Press Modify functional key to change the date/time.Press the WX functional
16 17Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source