The temperature has an effect on pH. The calibration buffer solutions are affected by temperature
changes to a lesser degree than normal solutions. During calibration the instrument will
automatically calibrate to the pH value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25 ºC.
PMETSREFFUBHp
CºFº10.400.603.8
023 10.421.684.8
514 00.490.644.8
0105 00.460.614.8
5195 00.440.673.8
0286 00.420.633.8
5277 10.400.603.8
0
368 20.499.572.8
5359 30.489.542.8
04401 40.479.512.8
pH BUFFER TEMPERATURE DEPENDENCE
Notes: To clear a previous calibration and to return to the default value, press Clear at any
time after entering calibration mode. The “Calibration cleared” message will be
shown for a few seconds on the display. If Clear is invoked during the first
calibration point the instrument returns to the measurement mode.
The Clear key is displayed only if a previous calibration exists.