Hanna-instruments HI 84431 Instrukcja Użytkownika Strona 20

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 19
20
The
(unstable measurement) symbol will be shown on the display until the reading becomes stable.
• When the reading is stable and close to the selected buffer, the
(unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
Press ESC to exit calibration.
Notes: To clear a previous calibration and to return to the default value, press Clear at any
time after entering calibration mode. The “Calibration cleared” message will be
shown for a few seconds on the display. If Clear is invoked during the first calibration
point the instrument returns to the measurement mode.
• The Clear key is displayed only if a previous calibration exists.
Point 2 calibration
• The calibrated value will be shown on the display and the second expected buffer value will
be displayed.
• Remove the electrode holder with electrodes from the top of the beaker.
Place the second buffer into beaker and place in beaker holder. Rinse the electrodes in a
beaker containing the second buffer rinsing solution.
• Place the electrode holder (with electrodes) on the top of the beaker, lock cap by turning.
• If necessary press the ARROW keys in order to select a different buffer value.
The
(unstable measurement) symbol will be shown on the display until the reading
becomes stable.
• When the reading is stable and close to the selected buffer, the
(unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
The calibrated value will be shown on the display and the third expected buffer value will be
automatically selected.
Przeglądanie stron 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 43 44

Komentarze do niniejszej Instrukcji

Brak uwag