Setup/Installation
pH calibration mode.
When the reading has stabilized, press
CONFIRM to accept the calibration or
CLEAR to restore the default calibration.
The “Wrong Bu er” message is displayed
when the pH value is outside of the
acceptable range. Clean the electrode by
following the cleaning procedure and/or
check the bu er value before continuing
the pH calibration. Press the ESC key to
exit pH calibration mode.
This screen is displayed when the
instrument is in titration mode. Press
START to begin a titration, METER to
enter pH meter mode or PRIME to enter
into the prime function.
The titration result, expressed as
concentration in selected unit, is displayed
automatically at the end of the titration.
Press RESTART to start another titration
or ESC to return to the main screen.
pH
Calibration
Messages
Titration
Messages