1QUICK START GUIDEHI 904KARL FISCHERCOULOMETRIC TITRATORRevision 1.0www.hannainst.com

Strona 2

10QUICK START GUIDEBEFORE PERFORMING THE FIRST TITRATIONSetup the Titrator• Make sure that all of the titrator assemblies are properly installed (see

Strona 3

7-6AUXILIARY FUNCTIONS7.3.2 Review Available ReportsUp to 100 reports can be saved on the titrator. To view one of the saved reports highlight areport

Strona 4

7-7AUXILIARY FUNCTIONSCompany Name:Allows the company name to be recorded in each report.Operator Name:Allows the operator name to be recorded in each

Strona 5

7-8AUXILIARY FUNCTIONS7.3.5 Setup Titration ReportCustomize a unique report to record the titration results. An asterisk means that it will beincluded

Strona 6

8-1MAINTENANCE, PERIPHERALSChapter 8. Contents8 MAINTENANCE, PERIPHERALS ... 8-38.1 Generato

Strona 7

8-2MAINTENANCE, PERIPHERALS

Strona 8

8-3MAINTENANCE, PERIPHERALS8 MAINTENANCE, PERIPHERALS8.1 Generator Electrode MaintenanceCaution: Never heat generator electrodes over 50ºC when drying

Strona 9

8-4MAINTENANCE, PERIPHERALSIf these signs are observed, the electrode pins may be dirty. Rinse the electrode with asolvent that is appropriate for the

Strona 10

8-5MAINTENANCE, PERIPHERALS5. Wipe joint grease off of the ground-glass joints with a clean, dry cloth or tissue.6. Dry the glass adapter in a drying

Strona 11

8-6MAINTENANCE, PERIPHERALS8.5.2 Connecting an External PC KeyboardThis connection allows you to use an external PS/2 PC Keyboard in addition to titra

Strona 12

8-7MAINTENANCE, PERIPHERALSThe correspondence between the Titrator’s Keypad and the United States 101-type externalkeyboar are:External PC Keyboard(Un

Strona 13

11QUICK START GUIDEFill Titration Vessel with ReagentThe titration vessel must be filled with reagent up to the MIN marker (about 75 mL):• Lower the r

Strona 14

8-8MAINTENANCE, PERIPHERALS8.5.3 Connecting a PrinterA variety of parallel printers can be connected to the parallel port of the titrator using astand

Strona 15

8-9MAINTENANCE, PERIPHERALS8.5.4 Connecting to a ComputerThe titrator can be connected to a computer using a USB cable. HI 900 PC applicationneeds to

Strona 16

8-10MAINTENANCE, PERIPHERALS

Strona 17 - INSTRUCTION MANUAL

9-1METHODS OPTIMIZATIONChapter 9. Contents9 OPTIMIZATION ... 9-39.1

Strona 18 - Hanna Instruments

9-2METHODS OPTIMIZATION9.2.5 Sample Preparation Techniques ... 9-99.2.5.1 Dilutions ..

Strona 19 - TABLE OF CONTENTS

9-3METHODS OPTIMIZATION9 OPTIMIZATION9.1 Titration SettingsThis section provides the descriptions of critical titration parameters necessary for an an

Strona 20

9-4METHODS OPTIMIZATIONnoitaziraloPtnerruC1 ΑµΑµ2 Αµ5 Αµ01tniopdnElaitnetoPVm09ot03 Vm031ot09 Vm081ot041 Vm003ot002The table that follows correlates e

Strona 21 - INTRODUCTION

9-5METHODS OPTIMIZATION9.1.2 Termination ParametersHI 904 provides a choice of three criteria by which a titration can be considered to havereached an

Strona 22

9-6METHODS OPTIMIZATION9.1.2.2.2 Absolute Drift StopUnder this criteria, a titration reaches an endpoint successfully when the drift falls below apred

Strona 23 - Chapter 2. Contents

9-7METHODS OPTIMIZATION9.2 The Sample9.2.1 Proper Sampling ProcedureProper sampling is essential for accurately determining the water content of bulk

Strona 24

12QUICK START GUIDELiquid Samples:• Samples with low viscosity will be added using a syringe with needle (injectionthrough the septum).• Weigh the syr

Strona 25 - 2.1 Unpacking

9-8METHODS OPTIMIZATION9.2.4 Liquid SamplesThe water contained in liquid samples must be available to react with the KF reagent. It isimportant to sel

Strona 26 - 2.2 Safety Measures

9-9METHODS OPTIMIZATION9.2.5 Sample Preparation TechniquesWhile many samples can be introduced directly into the titration vessel (see section 6.5Samp

Strona 27 - 2.3 Installation

9-10METHODS OPTIMIZATIONAn outline of a general procedure follows:1. Determine the mass of an extraction bottle or flask equipped with a septum.2. Add

Strona 28 - 2.3.3 Titrator Left-side View

9-11METHODS OPTIMIZATIONThe heating temperature is sample specific and can be found in applicable standard methods.The temperatures are chosen to be a

Strona 29 - 2.3.4 Titrator Assembly

9-12METHODS OPTIMIZATION

Strona 30

A3-2APPENDIX 3

Strona 31 - 2.3.4.3 Titration Vessel

A3-6APPENDIX 3MAN HI904 09/13

Strona 32

A1-1APPENDIX 1Appendix 1. ContentsA1 TECHNICAL SPECIFICATIONS ... A1-3

Strona 33

A1-2APPENDIX 1

Strona 34

A1-3APPENDIX 1A1 TECHNICAL SPECIFICATIONSRange 1 ppm to 5%Resolution 0.1 ppmResult Units %, ppm, mg/g, µg/g, mg, µg, mg/mL, µg/mL, ppt,mgBr/100g, gBr/

Strona 35 - Caution:

13QUICK START GUIDEUnderstanding the Displayed InformationDuring a titration, the following screen is displayed:Viewing Graph During TitrationPress

Strona 36

A1-4APPENDIX 1Reagent Handling SystemType Sealed system with integrated diaphragm air pumpDesiccant Type Molecular SievesBottle Thread Type GL-45Glass

Strona 37 - USER INTERFACE

A2-1APPENDIX 2Appendix 2. Recommended ReagentsA2 RECOMMENDED REAGENTS ... A2-3A2.1 Reage

Strona 38

A2-2APPENDIX 2

Strona 39

A2-3APPENDIX 2A2 RECOMMENDED REAGENTSA2.1 Reagents for Generators with DiaphragmSigma-Aldrich® HYDRANAL®34836 Coulomat AG (anolyte) 34840 Coulomat CG

Strona 41 - 3.2.2 Display

A3-1APPENDIX 3Appendix 3. AccessoriesA3 TITRATOR COMPONENTS ... A3-3

Strona 43 - Process Screen

A3-3APPENDIX 3HI 900561Titration Vessel(Glass only)HI 76330Detector ElectrodeHI 900511Generator Electrodewith DiaphragmHI 900512Generator Electrodewit

Strona 44

A3-4APPENDIX 3HI 900568Reagent ExchangeAdapterHI 900537Bottle Top Assembly(with molecular sieves)HI 900538Desiccant Cartridge forReagent/Waste Bottles

Strona 45

A3-5APPENDIX 3HI 900804Manual (English) forHI 904HI 900900UPC Application on USBFlash DriveHI 920013USB CableHI 900567Septum Kit (5 pcs)HI 900543Glass

Strona 46

14QUICK START GUIDE• The Review Result screen will be displayed.• Use the and keys to display information related to the last titration performed.

Strona 47 - Chapter 4. Contents

A3-6APPENDIX 3MAN HI904 09/13

Strona 48

1GENERAL APPLICATIONS BROCHUREHI 904KARL FISCHERCOULOMETRIC TITRATORwww.hannainst.com

Strona 50

Method ID: HI9001ENTitrator Validation with 1.0 mg/g Water StandardTitrator Application Methods: General Applications – r. 1.0Description:Method for

Strona 52

Method ID: HI9301ENMoisture Determination in Solventfor external dissolution or extractionTitrator Application Methods: General Applications – r. 1.0D

Strona 54

Method ID: HI9901ENBromine Index of Aromatic HydrocarbonsAdaptation of ASTM D1492-08Titrator Application Methods: General Applications – r. 1.0Descrip

Strona 55

Method ID: HI9901ENBromine Index of Aromatic HydrocarbonsAdaptation of ASTM D1492-08Titrator Application Methods: General Applications – r. 1.0Results

Strona 56 - Balance Configuration

1TITRATION THEORYTITRATION THEORYPrinciplesHI 904KARL FISCHERCOULUMETRIC TITRATORRevision 1.0www.hannainst.com

Strona 57

15QUICK START GUIDE• Transferring a report file will automatically transfer the corresponding log file andtitration graph BMP file (if applicable).• P

Strona 58

2TITRATION THEORY

Strona 59 - Chapter 5. Contents

3TITRATION THEORYContents1 GENERAL REVIEW OF TITRATION THEORY ... 51.1 Introduction to Titrations ...

Strona 60

4TITRATION THEORY4.3.2.2 Dispensing Errors... 224.3.3 Chem

Strona 61 - 5 METHODS

5TITRATION THEORY1 GENERAL REVIEW OF TITRATION THEORY1.1 Introduction to TitrationsA titration is a quantitative, volumetric procedure used in analyti

Strona 62 - 5.2 Standard Methods

6TITRATION THEORY1.3 Advantages and Disadvantages of TitrationsSome advantages of titrations, as an analytical technique, are:• More precise results t

Strona 63 - 5.3 User Methods

7TITRATION THEORYFigure 1.2 TYPES OF TITRATIONS2.1 Titrations According to The Measurement Method2.1.1 Amperometric TitrationsAn amperometric titratio

Strona 64 - 5.4 View / Modify Method

8TITRATION THEORYPotentiometry is the measurement of a potential under conditions of zero current flow. Themeasured potential can then be used to dete

Strona 65 - 5.5 Method Options

9TITRATION THEORY2.2 Titrations According to The Reaction Type2.2.1 Karl Fischer TitrationsThis method is based on a well-defined chemical reaction be

Strona 66 - 5.5.3 Method Type

10TITRATION THEORYof reaction decreases and titration endpoint become increasingly difficult to reach. If the pHexceeds 8, side reactions begin to occ

Strona 67 - 5.5.5 Stirring Speed

11TITRATION THEORYThe amount of current that passes through the generator can easily and accurately be measuredby the electronics of the titrator. The

Strona 68 - 5.5.7 Drift Entry

16QUICK START GUIDEQS 90408/13 Nr Titr Water[µg] mV Time 0 0.0 385.7 00:00:00 1 0.0 385.7 00:00:01 2

Strona 69 - 5.5.8 Reagent

12TITRATION THEORY2.2.1.4 Visual Indication of Karl Fischer TitrationsVisual methods, originally used by Karl Fischer, are limited in application, req

Strona 70 - 5.5.9 Sample Parameters

13TITRATION THEORY1502002503003504004500 0.2 0.4 0.6 0.8 1Potential (mV)Titrant Volume (mL)Karl Fischer Titration, Bivoltametric Indication vs. Titran

Strona 71 - Normal Determination

14TITRATION THEORYpClFigure 3.Figure 4.Figure 3 shows a traditional titration curve. The curve is obtained by plotting the pH valueagainst the volume

Strona 72 - Sample Type: Volume

15TITRATION THEORYFigure 5.2.2.4 Complexometric TitrationsA complex is a species where a central metal ion is covalently bonded to one or moreelectron

Strona 73 - External Dissolution

16TITRATION THEORYTitration of AcidsWeak acids with pKa’s up to about 11 can be titrated in non-aqueous solvents. These includecarboxylic acids, enols

Strona 74 - 5.5.10 Control Parameters

17TITRATION THEORYFigure 7.Figure 6.Visual indicators such as Ferrion are also available. The oxidized and reduced form of theindicator will have diff

Strona 75 - Standby Duration

18TITRATION THEORYABFigure 8.Potential, glass electrode, mVC

Strona 76

19TITRATION THEORY3 INTRODUCTION TO TITRATION APPARATUS AND TYPICALTITRATION PROCEDURE3.1 Manual TitrationApparatus required for manual titration incl

Strona 77 - 1 Reading

20TITRATION THEORY3.2 Automatic TitrationAutomatic titrators are high-precision analytical instruments that deliver the titrant, monitorthe physical c

Strona 78 - 5.5.11 Termination Parameters

21TITRATION THEORY4 TITRATION RESULTS4.1 AccuracyThe factors most critical to achieving accurate results with the HI 900 titration systems arethe conc

Strona 79 - Relative Drift

1INSTRUCTION MANUALHI 904KARL FISCHERCOULOMETRIC TITRATORRevision 1.00www.hannainst.com

Strona 80 - 5.6 Printing

22TITRATION THEORY4.3.2.2 Dispensing ErrorsIncorrect dispensing due to:• Dead valve volume and leaking valve;• Inaccuracy in motor drive and gear lash

Strona 81 - 5.7 Bromine Index

23TITRATION THEORY       

Strona 82

24TITRATION THEORY    

Strona 83 - TITRATION

25TITRATION THEORY        

Strona 84

26TITRATION THEORY       

Strona 85

27TITRATION THEORY  

Strona 86

28TITRATION THEORY6 GLOSSARYAcidA chemical species that can donate one or more protons (hydrogen ions).Acid-Base TitrationStoichiometric neutralizatio

Strona 87

29TITRATION THEORYEnd pointThe point where a titration is stopped because a physical change in the solution hasindicated a completed titration. Titrat

Strona 88

30TITRATION THEORYNeutralizationA chemical reaction where an acid and a base react to form a neutral salt and water.Non-aqueousA solution that does no

Strona 89

31TITRATION THEORYSpectrophotometric TitrationA titration in which the end point is marked by a change in the color and/or colorintensity.Stoichiometr

Strona 90

2Dear customer,Congratulations on choosing a Hanna Instruments product.Please read this instruction manual carefully before using the instrument. This

Strona 91

32TITRATION THEORYTitr THEORY 90407/13

Strona 92

3TABLE OF CONTENTSChapter 1. INTRODUCTIONChapter 2. SETUPChapter 3. USER INTERFACEChapter 4. GENERAL OPTIONSChapter 5. METHODSChapter 6. TITRATIONChap

Strona 93

2QUICK START GUIDEDear customer,Congratulations on choosing a Hanna Instruments Product.This guide has been written for the HI 904 Karl Fischer Coulom

Strona 95

1-1INTRODUCTION1 INTRODUCTIONThe HI 904 is an automatic coulometric Karl Fischer titrator with high accuracy, great flexibilityand repeatability.The t

Strona 96

1-2INTRODUCTION

Strona 97

2-1SETUPChapter 2. Contents2 SETUP ... 2-32.1 Unpacki

Strona 98

2-2SETUP

Strona 99

2-3SETUP2 SETUP2.1 UnpackingThe titrator and the accessories are shipped in a single box containing:ITEM QUANTITY1 Titrator ...

Strona 100 - AUXILIARY FUNCTIONS

2-4SETUP14 Power Adapter ... 1 pc.15 USB Cable ...

Strona 101

2-5SETUP2.3 Installation2.3.1 Titrator Top View

Strona 102

2-6SETUP2.3.2 Titrator Rear View2.3.3 Titrator Left-side View

Strona 103 - MAINTENANCE, PERIPHERALS

2-7SETUP2.3.4 Titrator AssemblyNote: Assembly operations must be completed before connecting the titrator to the power supply!2.3.4.1 Connecting the A

Strona 104

3QUICK START GUIDEContentsINTRODUCTION ...5SAFE

Strona 105

2-8SETUP2.3.4.2 Connecting the Reagent Adapter HolderThe Reagent Adapter Holder is a convenient dock for the Reagent Exchange Adapter when itis not i

Strona 106

2-9SETUPFigure 32.3.4.3 Titration VesselThe titration takes place in a sealed titration vessel. The titration vessel can also be referredto as a react

Strona 107

2-10SETUP2.3.4.3.1 Titration Vessel ComponentsWarning: Always apply a thin layer of joint grease (supplied) to ground-glass joints beforeattaching. I

Strona 108

2-11SETUP Reagent Exchange PortThe HI 904 titration vessel can be connected to the reagent and waste bottles using theReagent Exchange port and the su

Strona 109

2-12SETUP2.3.4.4 Electrical Connections• Connect the KF generator electrode to the 5-pin connector (C) using the supplied cable.• Connect the KF detec

Strona 110

2-13SETUP2.3.5 Reagent, Waste Bottle AssemblyThe bottle top assemblies are equipped with desiccant cartridges containing molecular sieves,which ensure

Strona 111

2-14SETUP2.3.6 Reagent Exchange AdapterThe Reagent Exchange Adapter is used to connect reagent and waste bottles to the titrationvessel. The adapter c

Strona 112

3-1USER INTERFACEChapter 3. Contents3 USER INTERFACE... 3-33.1 Start U

Strona 114

3-3USER INTERFACE3 USER INTERFACE3.1 Start UpOnce the instrument is assembled and installed, follow the steps below to start the titrator:• Connect th

Strona 115

4QUICK START GUIDE

Strona 116

3-4USER INTERFACE3.2 DescriptionThis chapter describes the basic principles of navigation through the user interface, selectingfields and entering val

Strona 117

3-5USER INTERFACE3.2.1.3 Arrow KeysThese keys have the following functions:• Move the on-screen cursor.• Increase and decrease the stirrer speed and o

Strona 118

3-6USER INTERFACE3.2.3 The Idle ScreenAfter start up and initialization, the first screen displayed is the Idle Screen.Idle Screen fields:Method name:

Strona 119

3-7USER INTERFACE3.2.4 The Process ScreenWhen the user presses while in Idle Screen, all titration related processes are started.The titrator displa

Strona 120

3-8USER INTERFACE3.3.1 Selecting an OptionTo select an option, simply press the option key below thevirtual key. For example, to access the Method Opt

Strona 121

3-9USER INTERFACE3.3.4 Saving ModificationsThe Saving Method screen allows the user to save themodifications. To exit from Saving Method screen withou

Strona 122

3-10USER INTERFACE

Strona 123

4-1GENERAL OPTIONSChapter 4. Contents4 GENERAL OPTIONS ... 4-34.1 Date and

Strona 124

4-2GENERAL OPTIONS

Strona 125 - APPENDIX 3

4-3GENERAL OPTIONS4 GENERAL OPTIONSThe General Options screen gives access to options that are not directly related to thetitration process. To access

Strona 126 - 09/13

5QUICK START GUIDEINTRODUCTIONThe HI 904 Karl Fischer Coulometric titrator is extremely flexible, capable of performing awide variety of highly accura

Strona 127 - APPENDIX 1

4-4GENERAL OPTIONS4.2 Display SettingsThis screen allows the user to customize the viewing features of the display.Option Keys:Increases the backlight

Strona 128

4-5GENERAL OPTIONS4.3 BeeperThis screen allows the user to turn the Beeper On (Enable) or Off (Disable).The beeper will sound after a titration is com

Strona 129

4-6GENERAL OPTIONS4.5 LanguageSelect an available language.4.6 Reagent Exchange ReminderCoulometric Karl Fischer reagents have a limited titration ca

Strona 130

4-7GENERAL OPTIONS4.7 Save Files to USB Storage DeviceThis option allows the user to save files from titrator to a USB storage device.On the titrator,

Strona 131 - APPENDIX 2

4-8GENERAL OPTIONS4.8 Restore Files from USB Storage DeviceThis screen allows the user to transfer files from the USB storage device to the titrator.I

Strona 132

4-9GENERAL OPTIONS4.9 USB Link with PCThe USB Link feature is useful to transfer methods/reports directly to/from a PC. To use thisfeature, connect th

Strona 133

4-10GENERAL OPTIONSDisconnect the KF electrode, then connect the HI 900940 calibration key to the electrodeinput (BNC connector).Depending on which pa

Strona 134

4-11GENERAL OPTIONSConfigure the settings on the titrator Balance Configuration menu to match the settings foryour particular balance (baud rate, data

Strona 135

4-12GENERAL OPTIONS4.13 Reset to Default SettingsThis option restores the manufacturer settings.Note: Please be careful!!! This will also delete all t

Strona 136

5-1METHODSChapter 5. Contents5 METHODS ... 5 - 35.1 Select

Strona 137

6QUICK START GUIDESAFETY MEASURESThe following safety measures must be followed:1. Never connect or disconnect the pump assemblies with the titrator t

Strona 138

5-2METHODS5.5.10 Control Parameters ... 5 - 165.5.10.1 Titration Speed

Strona 139

5-3METHODS5 METHODSAll of the parameters required to complete an analysis are grouped into a method.The titrator is supplied with a pack of standard m

Strona 140

5-4METHODS5.2 Standard MethodsThe standard methods were developed for the most common types of analysis. Also, thestandard methods can be used as a mo

Strona 141 - GENERAL APPLICATIONS BROCHURE

5-5METHODS5.2.3 Restoring the Standard Methods to the Manufacturer SettingsYou can restore the standard method to the manufacturer setting by highligh

Strona 142

5-6METHODSNote: Only a limited number of user methods can be generated. The titrator can hold 100methods (standard and user). When it is reached, a wa

Strona 143 - Method ID: HI9001EN

5-7METHODSSave method:After making modifications, press and select Save Method to keep the changes.5.5 Method Options5.5.1 Naming the User MethodTh

Strona 144

5-8METHODS5.5.2 Method RevisionThis option allows you to enter a string representing the current method revision. Therevision string format should be

Strona 145 - Method ID: HI9301EN

5-9METHODS5.5.4 Pre-Analysis Stir TimeTo avoid erroneous results or unreachable endpoints when analyzing samples with limitedsolubility, the sample mu

Strona 146

5-10METHODS5.5.6 Stirbar TypeAllows the user to edit the stirbar description.5.5.7 Drift EntryAllows the user to choose the drift entry mode that is u

Strona 147 - Method ID: HI9901EN

5-11METHODS5.5.8 ReagentThe user can enter a name for the reagent (up to 15 characters). Use the arrow keys tonavigate through the character table; pr

Strona 148

7QUICK START GUIDETITRATOR CONNECTIONSFront ViewRear View

Strona 149 - Revision 1.0

5-12METHODS5.5.9 Sample ParametersThis screen allows the user to access and configure the specific sample parameters.5.5.9.1 Sample DeterminationThis

Strona 150

5-13METHODS5.5.9.2 Sample NameThis screen allows the user to enter a name for the sample (up to 14 characters). Use thearrow keys to navigate through

Strona 151

5-14METHODS5.5.9.4 Sample SizeThis option allows the user to enter the sample size. For External/Dissolution, enter the sizeof the aliquot taken from

Strona 152

5-15METHODS5.5.9.6 External Solvent Size ( External Dissolution/Extraction determination modeonly)Enter the mass of the solvent used for external diss

Strona 153

5-16METHODS5.5.10 Control ParametersThe user can access and edit the parameters related to the titration.5.5.10.1 Titration SpeedChoose the desired ti

Strona 154

5-17METHODS5.5.10.2 Standby ModeWhen enabling this option, the titrator will automatically revert to Standby mode after atitration is completed.See al

Strona 155 - 2.1.1 Amperometric Titrations

5-18METHODS5.5.10.4 Imposed CurrentThe HI 904 uses a bivoltametric indicating electrode system. During a titration, the titratormonitors the voltage r

Strona 156

5-19METHODS5.5.10.6 Generator Current ModeThis option allows the user to select the generator electrode current mode:Auto:The titrator will dynamicall

Strona 157 - 2.2.1 Karl Fischer Titrations

5-20METHODS5.5.11.1 Maximum DurationSpecify the maximum time a titration is allowed to run. Once this point is reached, thetitration will be terminate

Strona 158 - ) = 9.336 · 10

5-21METHODS5.5.11.3 Termination CriterionThis screen allows the user to set the titration termination criterion.mV End PointThe titration is terminate

Strona 159

8QUICK START GUIDEUSER INTERFACEKeypadThe titrators have their own keypad with 29 keys grouped in four categories, as follows:DisplayThe titrators hav

Strona 160

5-22METHODS5.5.12 Result UnitThe titrator provides the results based on the selected units.5.6 PrintingTo print method parameters, press from the ma

Strona 161 - 2.2.2 Acid-Base Titrations

5-23METHODS5.7 Bromine IndexThe HI 904 is capable of performing Bromine Index/Bromine Number determinations. Bro-mine Index/Number is a measure of th

Strona 162

5-24METHODS

Strona 163

6-1TITRATIONChapter 6. Contents6 TITRATION ... 6-36.1 Idle ..

Strona 165 - 2.3.1 Back Titrations

6-3TITRATION6 TITRATION6.1 IdleThe titrator first enters Idle mode when it is switched on. All of the HI 904’s software featuresand adjustable paramet

Strona 166

6-4TITRATIONDuring the pre-titration, the user cannot change the currently selected method or access themethod parameters.Note: If the pre-titration l

Strona 167

6-5TITRATION6.4 StandbyAfter the drift rate has been determined, the HI 904 moves into Standby mode. In standbymode, the dryness of the titration vess

Strona 168

6-6TITRATION6.5 Sample AnalysisWhile in Standby mode, press .Note: If the drift value is zero a warning message appears to inform the user that theso

Strona 169 - 4.3.1 Sampling Errors

6-7TITRATIONManual EntryFollow the steps below:1. Attach a long needle (approximately 6 in. for best control) to a syringe that is largeenough to hold

Strona 170

9QUICK START GUIDEHOW TO SELECT YOUR LANGUAGETo change the language, press from the mainscreen. Highlight the Language option and thenpress . Using

Strona 171

6-8TITRATIONAutomatic Mass Acquisition from Analytical BalanceThe sample size can be automatically acquired from the balance when connected to thetitr

Strona 172

6-9TITRATIONThe titrator returns to the previous screen and the sample size is automatically updated.Now the analysis can be started.Start AnalysisPre

Strona 173

6-10TITRATIONViewing the Titration CurveDuring a titration, the titration curve can be displayed on the Titration Graph screen, bypressing . The titr

Strona 174 - 5.2.3 Blank Titration

6-11TITRATIONAveraging Sample Analysis ResultsBy pressing , the Sample Analysis History, Averaging and Statistical Data can beviewed.Use the and

Strona 176

7-1AUXILIARY FUNCTIONSChapter 7. Contents7. AUXILIARY FUNCTIONS ... 7-37.1 Air Pu

Strona 177

7-2AUXILIARY FUNCTIONS

Strona 178

7-3AUXILIARY FUNCTIONS7 AUXILIARY FUNCTIONS7.1 Air PumpThe air pump is used to add or remove the reagent in the titration vessel without exposureto at

Strona 179

7-4AUXILIARY FUNCTIONS7.2 StirrerNote: When custom stirrer is selected (see Section 4.4, Stirrer in General Options chapter),the commands related to t

Strona 180

7-5AUXILIARY FUNCTIONS7.3 ResultsTo access the “Data Parameters” screen, press button.From the Data Parameters screen you can access the following o