Hanna-instruments HI 93710 Instrukcja Użytkownika Strona 46

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 52
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 45
46
pp
pp
p
H MEASUREMENTH MEASUREMENT
H MEASUREMENTH MEASUREMENT
H MEASUREMENT
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will appear on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
Remove the cuvet and add 5
drops of the HI 93710 Phenol
Red Indicator. Replace the cap
and swirl the solution.
Reinsert the cuvet into the instrument.
Press the READ key and “SIP” will appear on the display during
measurement.
The instrument directly displays the pH measured value on the
Liquid Crystal Display.
ZERO
pH
10 mL
x 5
Przeglądanie stron 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 52

Komentarze do niniejszej Instrukcji

Brak uwag