Hanna-instruments HI 95721 Instrukcja Użytkownika Strona 5

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 10
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 4
5
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Precision
±0.01 mg/L @ 1.50 mg/L
Typical EMC Deviation ±0.01 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of the USEPA method 315B and Standard Method
3500-Fe B. The reaction between iron and the phenanthroline
reagent causes an orange tint in the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off
After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity/test
HI 93721-0 Iron HR Reagent 1 packet
Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 1.50
mg/L Iron and a representative lot of reagent, an
operator obtained with a single instrument a
standard deviation of
0.01 mg/L.
SPECIFICATIONS
PRECISION AND ACCURACY
16
CALIBRATION PROCEDURE
Turn the meter on by pressing ON/OFF.
When the LCD displays “721”, it is ready.
Enter the calibration mode by holding CAL CHECK
for three seconds.
The date of the last calibration appears (e.g.:
month “01”, day “08”). “F.CAL” means that the
factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION, see
instructions page 18.
Place the CAL CHECK™ Standard Cuvet A into
the holder and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
After a few seconds the display will show “-0.0-”.
The meter is now zeroed and ready for calibration.
Remove the cuvet.
Place the CAL CHECK™ Standard HI 95721-11
Cuvet B into the holder and ensure that the notch
on the cap is positioned securely into the groove.
Note: it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna CAL
CHECK™ Standards, otherwise erroneous results will be obtained.
or
CALIBRATION PROCEDURE
Przeglądanie stron 4
1 2 3 4 5 6 7 8 9 10

Komentarze do niniejszej Instrukcji

Brak uwag