This Instrument is in
Compliance with the CE Directives
Instruction Manual
HI 95724C
Free & Total Chlorine
ISM
24
MAN95724C 01/04
www.hannainst.com
Przeglądanie stron 0
1 2 3 4 5 6 ... 11 12

Podsumowanie treści

Strona 1 - Free & Total Chlorine

1www.hannainst.comThis Instrument is inCompliance with the CE DirectivesInstruction ManualHI 95724CFree & Total Chlorine ISM24MAN95724C 01/04www

Strona 2 - USER NOTES

10“Error”: the concentration of the calibration solution used is not correct.Repeat the calibration procedure with the right standard solution, andver

Strona 3 - HANNA LITERATURE

11GENERAL TIPS FOR AN ACCURATE MEASUREMENTThe instructions listed below should be carefully followed during testingto ensure best accuracy.• For a cor

Strona 4 - CE DECLARATION OF CONFORMITY

12MEASUREMENT PROCEDURE• Turn the meter on by pressing ON/OFF.• When the LCD displays “724”, it is ready.• Do not let the reacted sample stand too lon

Strona 5 - WARRANTY

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Strona 6 - ACCESSORIES

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occured during shipment, ple

Strona 7 - BATTERY REPLACEMENT

4The HI 95724 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru

Strona 8

5Range 0.00 to 5.00 mg/L of Cl2 (Free or Total)Resolution 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/LPrecision ±0.03 mg/L @ 1.00 mg/LT

Strona 9 - GUIDE TO DISPLAY CODES

6Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some o

Strona 10 - CALIBRATION PROCEDURE

7FUNCTIONAL DESCRIPTION1) Liquid Crystal Display2) Cuvet Holder3) Cuvet alignment indicator4) ON/OFF key5) ZERO key6) READ key7) CAL CHECK keyA microp

Strona 11 - VALIDATION PROCEDURE

8KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: press to zero the meter prior to measurement.• READ: press for making a measurement.

Strona 12 - MEASUREMENT PROCEDURE

9This prompt appears for 1 second each time the instrument is turned on.The parameter code “724” indicates that the meter is in a ready stateand zeroi