Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Alternatively, wait for one minute and
just press READ/ /TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display, de-
pending on the measurement phase.
The instrument directly displays
concentration in mg/L of free chlorine on
the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Chlorine Dioxide
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L
CaCO
3
, the color of the sample could disappear or develop only partially. To resolve this,
neutralize the sample with diluted HCl or NaOH.