Hanna-instruments HI 96718 Instrukcja Użytkownika Strona 14

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 24
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 13
14
Remove the cuvette.
Add the content of one packet of
HI 93718-0 Iodine reagent. Replace the
cap and shake gently for about 20 seconds
to dissolve most of the reagent.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 2 minutes and 30
seconds then just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
iodine on the LCD.
INTERFERENCES
Bromine
Chlorine
Ozone
Oxidized forms of chromium and manganese
Przeglądanie stron 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 23 24

Komentarze do niniejszej Instrukcji

Brak uwag