10Under range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam

Strona 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Strona 4 - ABBREVIATIONS

12• It is important that the sample does not contain any debris. This wouldcorrupt the reading.• Each time the cuvette is used, the cap must be tighte

Strona 5 - PRECISION AND ACCURACY

13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed

Strona 6 - PRINCIPLE OF OPERATION

14• Remove the blank and insert the sampleinto the instrument and ensure that thenotch on the cap is positioned securelyinto the groove.• Press and ho

Strona 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDUREUse the validation procedure to ensure that the instrument is properly calibrated.Warning: Do not validate the instrument with a

Strona 8

16• After a few seconds, the display will show“-0.0-”. The meter is now zeroed and readyfor validation.• Remove the cuvette.• Place the CAL CHECKTM St

Strona 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Strona 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed and readyfor calibration. The blinking “READ” asks forreading calibratio

Strona 11 - (c)(a) (b)

19• Press GLP/ to edit the desired month(01-12). If the key is kept pressed, themonth number is automatically increased.• Press GLP/ to edit the des

Strona 12 - MEASUREMENT PROCEDURE

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Strona 13

20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco

Strona 14

21• Press ZERO/CFM to restore the factorycalibration or press GLP/ again toabort factory calibration restore.• Press READ/ /TIMER to enter in thefact

Strona 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Strona 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Strona 17 - CALIBRATION PROCEDURE

24MAN96719 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Strona 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Strona 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96719 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Strona 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.11 mg/L ±5% of reading @ 25°CTypical EMC Deviation ±0.02 m

Strona 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Strona 22 - BATTERY REPLACEMENT

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette h

Strona 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or t

Strona 24 - (ROMANIA)

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs