The instructions listed below should be carefully followed during testing to ensure best accuracy.
Use matched square cuvettes, having 4 clear faces and 10
mm path length.
Do not touch the cuvette walls with hands.
Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil
or dirt. Wipe it thoroughly with HI 731318 or a lint-free
cloth prior to insertion.
It is important that the sample does not contain any debris.
This would corrupt the reading.
Samples should be completely free of air bubbles. If bubbles are present tap the cuvette on
the table to remove them.
It is recommended to take the sample of honey in the middle of the container.
It is recommended to take and analyze samples at room temperature.
Do not use the same cuvette with glycerol for more than 5-10 times to perform calibration
and dispose of it after 1 day has passed.
Never re-use honey sample cuvettes, dispose them after use.
Note: Honey samples exposed to direct sun light and/or air will darken over time. This is
caused by microorganisms.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
Over Range: A flashing value of the maximum concentration
indicates an over range condition. The concentration of the
sample is beyond the programmed range.
Cap error: Appears when external light enters in the analysis
cell. Assure that the light shield cap is present.
Cooling lamp: The instrument waits for the lamp to cool down.
d) other errors and warnings
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, the meter will
lock up. Change the battery and restart the meter.
Under range: A blinking “0” indicates that the sample absorbs
less light than the calibration reference. Check the procedure and
recalibrate the instrument.