The TDS value in aqueous solutions is directly proportional to the
conductivity. The ratio between the two parameters depends on the
solution.
HI 9811-5 model has a fixed conversion factor set to 0.5. This
means that 1 µS/cm is equal to 0.5 mg/L of TDS.
Accessories needed:
• Use HI 70031 (1413 µS/cm) EC calibration solution or
HI 70032 (1382 ppm, or mg/L) TDS calibration solution.
Note: The conversion between EC and TDS is made by a built-in
circuit, hence it is requested to calibrate the meter only in
EC or TDS range. The other range is thus automatically
calibrated.
PROCEDURE
• Pour approximately 4 cm (1½") of a conductivity calibration
solution (e.g. HI 70031) into a beaker. If possible, use plastic
beaker to minimize any EMC interference.
• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be
reached.
• Tap the probe on the bottom, then shake it lightly while rotating
to make sure no air bubbles remain trapped inside the probe.
• Press the µS/cm (or mg/L) key.
• Turn the EC/TDS calibration knob until
the display shows the EC or TDS reading
at 25°C.
EC/TDSEC/TDS
EC/TDSEC/TDS
EC/TDS
CALIBRATION CALIBRATION
CALIBRATION CALIBRATION
CALIBRATION
EC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOR
EC/TDS CONVERSION FACTOREC/TDS CONVERSION FACTOR
EC/TDS CONVERSION FACTOR