Strona 1 - HI 98140 • HI 98150

Instruction ManualHI 98140 • HI 98150Portable WaterproofpH / ORP Meterswith Smart Electrodes& Log-On-Demandwww.hannainst.com

Strona 2 - TABLE OF CONTENTS

1918Note: Press ALT and DEL. to escape without data deletion.When viewing through the logged data, the "NUL" message will bedisplayed when s

Strona 3 - FUNCTIONAL DESCRIPTION

2120GLP AND RS232All the GLP data can be retrieved (if the electrode is connected) froma PC through the RS232 communication feature. The calibration d

Strona 4 - SPECIFICATIONS

2322The commands available are as follows:MOD - to request the firmware code of the meter.GLP - to request the last calibration data.The meter answers

Strona 5 - SETUP MODE

2524TEMPERATURE-RESISTANCE CORRELATION FORHANNA pH SENSITIVE GLASSThe resistance of glass electrodes partially depends on the temperature.The lower th

Strona 6 - TAKING pH MEASUREMENTS

2726Replace solution in the protective cap with a few drops of HI 70300storage solution or, in its absence, filling solution (HI 7071 for singlejuncti

Strona 7 - TAKING ORP MEASUREMENTS

2928ACCESSORIESpH CALIBRATION SOLUTIONSHI 70004P pH 4.01 buffer solution, 20 mL sachet, 25 pcsHI 70007P pH 7.01 buffer solution, 20 mL sachet, 25 pcsH

Strona 8 - CALIBRATION ERROR MESSAGES

CE DECLARATION OF CONFORMITYWARRANTYAll Hanna Instruments meters are warranted for two yearsagainst defects in workmanship and materials when used for

Strona 9 - LOGGING FUNCTION

MAN98150R4 08/05SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 8857006

Strona 10 - TO DELETE LOGGED DATA

32Remove the instrument from the packing material and examine itcarefully to make sure that no damage has occurred during shipping.If there is any dam

Strona 11 - CALIBRATION ALARM TIME-OUT

54FUNCTIONAL DESCRIPTION1) Power adapter plug2) RS232 connector3) DIN connector for smart electrodes4) Liquid Crystal Display (LCD)5) ON/OFF key, to t

Strona 12 - DATA TRANSFER TO PC

76INITIAL PREPARATIONEach meter is supplied complete with batteries. Remove the backcover, unwrap the batteries and install them while paying attentio

Strona 13 - BATTERY REPLACEMENT

98PARAMETER SETTINGS• Once the parameter code (and passwordif needed) has been entered, the currentvalue of the selected parameter on theupper LCD and

Strona 14 - AND MAINTENANCE

1110Connect the pH electrode with the built-in tempera-ture sensor to the meter and press ON/OFF topower on the instrument. The meter automaticallyche

Strona 15 - ACCESSORIES

1312pH CALIBRATION PROCEDUREFor greatest accuracy, it is recommended that the instrument iscalibrated frequently. For a faster procedure, it is possib

Strona 16 - WARRANTY

151415. The LCD will display "Stor". The instrument checks the electrodeparameters and advises the user of abnormalities by "old probe&

Strona 17 - MAN98150R4 08/05

1716To store the current reading in memory press the LOG key while inmeasurement mode. The LCD will display "Stor" along with the "LOG&