Hanna-instruments HI 9912 Instrukcja Użytkownika Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 17
18
Please refer to the following chart for a more accurate pH calibra-
tion:
For instance, if the buffer temperature is 25°C (77°F), calibrate the
meter to read 4.01, 7.01 or 10.01 on the display.
If the buffer temperature is 20°C, calibrate it to 4.00, 7.03 or
10.06.
If the buffer temperature is 50°C, calibrate it to 4.06, 6.98 or 9.82.
pHpH
pHpH
pH
VALUES AT VARIOUS VALUES AT VARIOUS
VALUES AT VARIOUS VALUES AT VARIOUS
VALUES AT VARIOUS
TEMPERATURESTEMPERATURES
TEMPERATURESTEMPERATURES
TEMPERATURES
TEMP pH VALUES
°C °F 4.01 6.86 7.01 9.18 10.01
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
Przeglądanie stron 17