PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully
to make sure that no damage has occurred during shipping. If there is any
damage, notify your Dealer or the nearest Hanna Customer Service Center.
Each instrument is supplied with:
HI 1131B Glass-body Combination pH Electrode with 1 m (3.3')
Cable
HI 7662 Temperature Probe
HI 76404N Electrode Holder
pH 4.01 & 7.01 Buffer Solutions (20 mL each)
HI 7071 Electrolyte Solution
HI 700661 Cleaning Solution
12VDC Power Adapter
Instruction Manual
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.
The HI 2216 is state of the art, heavy-duty pH, mV, ISE meter designed
to provide laboratory results and accuracy under harsh industrial
conditions.
HI 2216 can also measure Oxidation Reduction Potential (ORP) in mV
range.
Relative mV feature is also provided.
pH measurements are compensated for temperature effect manually or
automatically with the HI 7662 temperature probe.
Up to a five-point pH calibration can be performed using seven standard
buffers. In addition, two custom buffers can be used during calibration.
Up to a two-point ISE calibration can be performed using five standard
solutions.
The GLP feature provides data consistency.
A calibration due alarm can be set to alert the user that too much time
elapsed since the last pH calibration.
Data can be stored in meter’s memory for later retrieval.
An USB connection ensures communication with a PC.
The meter’s memory can hold 200 manually logged points and 500 lot
logging points.