1HI 93414Turbidity and Free/Total ChlorineMeterInstruction Manualwww.hannainst.com

Strona 2 - WARRANTY

108) ON/OFF, press to turn the instrument ON/OFF. If no key is pressed for more than15 minutes, the instrument automatically shuts off.9) RANGE s, pre

Strona 3 - PRELIMINARY EXAMINATION

11DISPLAY DESCRIPTION1) Battery icon. When the instrument is powered by batteries, at the start of the instrument, theremaining battery life is displa

Strona 4 - GENERAL DESCRIPTION

12SPECIFICATIONSTurbidityRange 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTURange selection AutomaticallyResolution 0.01 NTU from 0.00 to 9.99 NTU;0.

Strona 5 - ABBREVIATIONS

13OtherLight Source Tungsten filament lampLamp life greater than 100,000 readingsDisplay 60 x 90mm LCD with backlightLOG Memory 200 recordsSerial Inte

Strona 6 - PRINCIPLE OF OPERATION

14GENERAL TIPS FOR AN ACCURATE MEASUREMENTCUVETTEThe cuvette is part of the optical system in all measurements. The light reaches the sample bypassing

Strona 7 - COLORIMETER

15CUVETTE OILING (TURBIDITY only)Warning: For colorimetric measurements the cuvette should be completely free of any trace of oil.Do not use the oilin

Strona 8

16In order to index a cuvette follow the next steps:• Fill the cuvette with high quality water (<0.1NTU) up to the 10 mL mark.• Clean and oil the c

Strona 9 - FUNCTIONAL DESCRIPTION

17In order to match multiple cuvettes follow the nextsteps:• Fill some cuvettes with high quality water(<0.1NTU) up to the 10 mL mark.• Clean and o

Strona 10

18• Open the instrument lid, slightly rotate the cuvette and take a new reading.• Repeat the last step for the second cuvette until the reading is wit

Strona 11

19APPLICATION OF VACUUMVacuum works by decreasing the atmospheric pressure. In this way the bubbles from the solutioncame out to the surface.Applicati

Strona 12 - SPECIFICATIONS

2TABLE OF CONTENTSWARRANTYWARRANTY ...

Strona 13

20HEATING THE SAMPLEUse of heat to remove air bubbles, although very effective in some cases, should be handled with carebecause it can alter the turb

Strona 14 - CUVETTE PREPARATION

21RANGE SELECTIONThe HI 93414 instrument has three measurement ranges:• Turbidity from 0.00 to 1000 NTU• Free chlorine from 0.00 to 5.00 mg/L• Total c

Strona 15 - INDEXING A CUVETTE

22MEASUREMENT PROCEDURETURBIDITY MEASUREMENTSTo take turbidity measurements, follow the nextsteps:• Turn the instrument ON by pressing ON/OFF.When das

Strona 16

23NORMAL MEASUREMENTThis type of measurement can be used for regular readings, when the sample is stable and normalaccuracy is required. In normal mea

Strona 17 - MATCHING MULTIPLE CUVETTES

24• Press READ/TIMER u to start the averagereading. The display will show blinking dashesand the icons for cuvette, detectors and lampwill appear duri

Strona 18

25FREE AND TOTAL CHLORINE MEASUREMENTTo take colorimetric measurements follow nextsteps:• Turn the instrument on by pressing ON/OFF.Assure that the co

Strona 19 - USE OF AN ULTRASONIC BATH

26SINGLE SAMPLE READ• Remove the cuvette from the instrument.• Remove the cap.• Add the content of one packet of the specifictest reagent, for:Free Ch

Strona 20 - HEATING THE SAMPLE

27MULTIPLE SAMPLES READ• Place the second cuvette with the reactedsample into the holder and ensure that themark on the glass is matched with the mark

Strona 21 - RANGE SELECTION

28CALIBRATION PROCEDURETURBIDIMETER CALIBRATIONHI 93414 has a powerful calibration function that compensates for lamp aging or changing. Thecalibratio

Strona 22 - MEASUREMENT PROCEDURE

29TWO-POINT CALIBRATION• Turn the instrument ON by pressing ON/OFF.If you are not in turbidity range, first select therange. If you are in turbidity r

Strona 23 - AVERAGED MEASUREMENT

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Strona 24 - RANGE AND UNITS

30• Close the lid and press READ/TIMER u. Thedisplay will show blinking dashes and the iconsfor cuvette, detectors and lamp will appearduring measurem

Strona 25 - ZEROING THE INSTRUMENT

31FOUR-POINT CALIBRATIONTo perform a four-point calibration, continue theprocedure with the following steps:• Remove the third standard cuvette.• Plac

Strona 26 - SINGLE SAMPLE READ

32• If the calculated calibration coefficients areoutside a certain range the “CAL Err” messageis displayed.CALIBRATION DELETIONHI 93414 is factory ca

Strona 27 - INTERFERENCES

33• Turn the instrument on by pressing ON/OFF.Make sure that the instrument is in the free ortotal chlorine range (the desired one).• Press RANGE s to

Strona 28 - TURBIDIMETER CALIBRATION

34To calibrate one range follow next steps:• Turn the instrument ON by pressing ON/OFF.Make sure that the instrument is in the free ortotal chlorine r

Strona 29 - TWO-POINT CALIBRATION

35• Press and hold CAL/CHECK for 3 seconds toenter calibration. The LCD will show “CAL”tag and the parameter for which the calibrationis performed.•

Strona 30 - THREE-POINT CALIBRATION

36To delete last calibration, follow next steps:• Enter the GLP feature by pressing RCL/GLP for3 seconds. The date of the last calibration willbe disp

Strona 31 - FOUR-POINT CALIBRATION

37LOGGINGLOGGINGThe log function is active only after a validmeasurement is obtained (no errors).• To log a value, press LOG/CFM when themeasurement r

Strona 32

38VIEW LOGGED DATAThe stored records can be viewed at any momentby pressing RCL/GLP. To return to normalmeasurement mode press RCL/GLP again.LOG SEARC

Strona 33

39• Measurement date in YYYY.MM.DD format.• Measurement time in hh:mm format.• Delete the last record panel (only for lastrecord).• Delete all records

Strona 34

4HI 93414 is a high accuracy, combined meter that benefits from Hanna’s years of experience asmanufacturer of analytical instruments.The HI 93414 succ

Strona 35

40• Press and hold RCL/GLP for 3 seconds toenter/exit GLP data consulting. Several functionsare available when in GLP menu.• Press READ/TIMER u to scr

Strona 36 - CALIBRATION ERROR MESSAGES

41• Third calibration point - only for turbidimeterrange (if available).• Fourth calibration point - only for turbidimeterrange (if available).• Delet

Strona 37

42SETUPSetup mode allows viewing and modifying the instrument parameters.The blinking “CAL” tag during setup mode suggest to press CAL/CHECK for param

Strona 38 - RECORD VIEWING

43SET BEEPERThe HI 93414 has a built in beeper that signals the tag read, the key press and the error conditions.The beeper can be selected to be ON o

Strona 39 - DELETE ALL RECORDS

44• To set the current date, press CAL/CHECK whenset date panel is displayed. The date format isYYYY.MM.DD. The last two digits of the yearvalue and “

Strona 40

45• Press the s or t keys to set the hour value.• Press LOG/CFM or READ/TIMER u to startediting the minutes. The minutes value willstart blinking.• Pr

Strona 41

46• To set the baud rate, press CAL/CHECK whenset baud rate panel is displayed. The parametervalue and “CFM” tag will start blinking.• Press the s or

Strona 42

47TAG INSTALLATIONThe tag is housed in a rugged metal that can withstand harsh environments. However, it is betterto protect the tag from direct rain.

Strona 43

48BATTERIES REPLACEMENTTo replace the batteries follow the next steps:• Press ON/OFF to turn OFF the instrument.• Open the batteries cover by pressing

Strona 44

49• Replace the cover and press it until it locks.• Turn the instrument ON.Warning: Replace batteries only in a non-hazardous area.USING AN AC ADAPTER

Strona 45

5TAG IDENTIFICATION SYSTEMHanna is the first manufacturer of analytical instruments that has decided to add the unique T.I.S.-Tag Identification Syste

Strona 46 - LCD BACKLIGHT

50ERROR CODESHI 93414 has a powerful diagnostic system. The common errors are detected and reported foreasy diagnostic and maintenance.RORRE NOITPIRCS

Strona 47 - LAMP REPLACEMENT

51RECOMMENDATIONS FOR USERSBefore using this product, make sure that it is entirely suitable for your specific application andfor the environment in w

Strona 48 - BATTERIES MANAGEMENT

52Printed in ROMANIAMAN93414 04/13Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for Cust

Strona 49 - PC INTERFACE

6TURBIDIMETERTurbidity of water is an optical property that causes light to be scattered and absorbed, rather thantransmitted. The scattering of the l

Strona 50 - ERROR CODES

7MEASUREMENT UNITSMany methods were used to measure turbidity over the years. The Jackson Candle Turbidimeter wasused to measure turbidity as Jackson

Strona 51 - ACCESSORIES

8Therefore, the concentration “c” can be calculated from the absorbance of the substance as the otherfactors are known.Photometric chemical analysis i

Strona 52 - MAN93414 04/13

9FUNCTIONAL DESCRIPTIONINSTRUMENT DESCRIPTION1) Cuvette Lid. Close the cuvette lid prior to start a measurement.2) Cuvette Holder. Insert the cuvette