Hanna-instruments HI 2300 Instrukcja Użytkownika Strona 30

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 29
30
PAR Requests the setup parameters setting.
The answer string contains:
Instrument ID (4 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
cell constant, with sign and decimal point (6 chars)
TDS factor, with sign and decimal point (5 chars)
TC coef, with sign and decimal point (5 chars)
Reference Temperature, with sign and decimal point
(5 chars)
Temperature Compensation mode (1 char)
NSLx Requests the number of logged samples.
x = E - request for EC range
N - request for NaCl range
T - request for TDS range
LODExxx: Request the xxx
th
EC record logged data.
LODNxxx: Request the xxx
th
NaCl record logged data.
LODTxxx: Request the xxx
th
TDS record logged data.
LODEALLxx:Request all EC Log on demand. See Note.
LODNALLxx:Request all NaCl Log on demand. See Note.
LODTALLxx:Request all TDS Log on demand. See Note.
Note: The data is sent in frames of 8 records each. “xx” represents the
number of the frame. The first frame is labeled 01.
Errors: “Err3” log on demand empty.
“Err4” requested set parameter is not available.
“Err5” command argument is wrong.
“Err6” requested range not available.
“Err7” meter in log mode.
“Err8” is sent if instrument is not in measurement mode.
“NAK” (21) character is sent when the instrument receives
an unknown or a corrupted command.
Note: E - request for EC range.
T - request for TDS range.
N - request for NaCl range.