Silica Low Range
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 705 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• One bottle of Reagent A
• Twelve packets of Reagent B
• Twelve packets of Reagent C
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 705
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
For more details about spare parts and accessories
see “Accessories”.
Technical specifications:
Range 0.00 to 2.00 ppm
Resolution 0.01 ppm
Accuracy ±0.03 ppm ±5% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 610 nm
Light Detector Silicon Photocell
Method Adaptation of the ASTM D859 method of heteropoly blue
method. The reaction between silica and reagents causes
a blue tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected]
or view our worldwide contact list at www.hannainst.com.
Przeglądanie stron 0
1 2

Podsumowanie treści

Strona 1 - Silica Low Range

Silica Low RangePreliminary examination:Please examine this product carefully. Make sure that the instrument is not damaged. Ifany damage occurred dur

Strona 2 - Battery management

10 mL• It is important that the sample does not contain any debris.• Whenever the cuvette is placed into the measurement cell, it must be dryoutside,

Komentarze do niniejszej Instrukcji

Brak uwag