Phosphate LR
PHOSPHATE LOW RANGE
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.01 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Ascorbic Acid method. The reaction between phosphate and the
reagent causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93713-0 Powder reagent 1 packet
REAGENT SETS
HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 124.
MEASUREMENT PROCEDURE
Select the
Phosphate LR
method using the procedure
described in the
Method Selection
section (see page 12).
Rinse, cap and shake the cuvette several times with
unreacted sample. Fill the cuvette with 10 mL of sample
up to the mark and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0- when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet of
HI 93713-0 reagent. Replace the cap and shake gently (for
about 2 minutes) until the powder is completely dissolved.
Phosphate HR
Add the content of one packet of HI 93717B-0 Phosphate
HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of phosphate (PO
4
3
¯).
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of phosphorus (P) and phosphorus pentoxide
(P
2
O
5
).
Press s or t to return to the measurement screen.
INTERFERENCES
Sulfide
Chloride above 150000 mg/L )
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L
10 mL