Hanna-instruments HI 83207-01 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Hanna-instruments HI 83207-01. Hanna Instruments HI 83207-01 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - HI 83207

1Instruction ManualHI 83207Multiparameter BenchPhotometerfor Industrial Wastewaterwww.hannainst.com

Strona 2 - TABLE OF CONTENTS

10• Pay attention to push the cuvette completely down in the holder and to align the white point on thecap to the indexing mark on the meter.• In orde

Strona 3 - GENERAL DESCRIPTION

11METHOD REFERENCE TABLEMethod Method Pagedescription1 Aluminum 172 Free Chlorine 193 Total Chlorine 224 Color of Water 255 Copper HR 276 Copper LR 29

Strona 4 - SPECIFICATIONS

12POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note: Alw

Strona 5

13DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measurem

Strona 6 - FUNCTIONAL DESCRIPTION

14In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that ch

Strona 7 - KEYPAD DESCRIPTION

15Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the W X functional keys to h

Strona 8

16HI 83207 offers an interactive contextual help mode that assists the user at any time.To access help screens press HELP.The instrument will display

Strona 9 - USING CUVETTES

17AluminumALUMINUMSPECIFICATIONSRange 0.00 to 1.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationL

Strona 10 - INTERFERENCES

18Aluminum• Add the content of one packet of HI 93712C-0 Bleachingpowder to one of the two cuvettes. Replace the cap and shakevigorously until complet

Strona 11 - HEALTH & SAFETY

19FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of

Strona 12 - OPERATIONAL GUIDE

2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will prov

Strona 13 - CHEMICAL FORM

20Free Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Strona 14 - SETUPSETUP

21• Reinsert the cuvette into the instrument.• Press Read to start the reading. The instrument displays the results in mg/L of free chlorine.INTERFER

Strona 15

22Total ChlorineTOTAL CHLORINESPECIFICATIONSRange 0.00 to 3.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg

Strona 16 - HELP MODEHELP MODE

23Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reag

Strona 17 - ALUMINUM

24• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor 2 minutes and 30 seconds and press Read. W

Strona 18

25Color of WaterCOLOR OF WATERSPECIFICATIONSRange 0 to 500 PCU (Platinum Cobalt Units)Resolution 1 PCUAccuracy ±10 PCU ±5% of reading at 25 °CTypical

Strona 19 - FREE CHLORINE

26Color of Water• Fill the second cuvette up to the mark with unfiltered sampleand replace the cap. This is the apparent color.• Filter 10 mL of sampl

Strona 20 - Free Chlorine

27Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/L

Strona 21

28Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Strona 22 - TOTAL CHLORINE

29Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of reading at 25 °CTypical EMC ±1 μg/LDeviationLig

Strona 23 - Total Chlorine

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Strona 24

30Copper LR• Remove the cuvette.• Add the content of one packet of HI 95747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Strona 25 - COLOR OF WATER

31FluorideFLUORIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationL

Strona 26 - Color of Water

32• Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, waitfor two minutes and press Zero. The display

Strona 27 - COPPER HIGH RANGE

33SPECIFICATIONSRange 0.0 to 20.0 mg/LResolution 0.1 mg/LAccuracy ±0.2 mg/L ±3% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source Tungsten

Strona 28

34• Add the content of one packet of HI 93709B-0 SodiumPeriodate reagent. Replace the cap and invert to mix withgently movements for 2 minutes.• Reins

Strona 29 - COPPER LOW RANGE

35Manganese LRMANGANESE LOW RANGESPECIFICATIONSRange 0 to 300 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±3% of reading at 25 °CTypical EMC ±1 μg/LDeviati

Strona 30

36Manganese LR• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagentsolution to each cuvette, replace the caps and invert gentlyto mix for about 30

Strona 31 - FLUORIDE

37Manganese LR• Insert the second cuvette with the reacted sample into the instrument.• Press Read to start the reading. The instrument displays the r

Strona 32

38MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationL

Strona 33 - MANGANESE HIGH RANGE

39• Fill one graduated mixing cylinder up to the 25 mLmark with the sample.• Add the content of one packet of HI 93730A-0molybdenum reagent, close the

Strona 34

4PRINCIPLE OF OPERATIONAbsorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.When a light beam cross

Strona 35 - MANGANESE LOW RANGE

40Molybdenum• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor five minutes and press Read. Whe

Strona 36 - Manganese LR

41Nickel HRSPECIFICATIONSRange 0.00 to 7.00 g/LResolution 0.01 g/LAccuracy ±0.07 ±4% of reading at 25 °CTypical EMC ±0.02 g/LDeviationLight Source Tun

Strona 37

42Nickel HR• Remove the cuvette and add the content of one packet ofHI 93726-0 reagent. Replace the cap and shake gently untilcompletely dissolved.• R

Strona 38 - MOLYBDENUM

43Nickel LRNICKEL LOW RANGESPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.001 mg/LAccuracy ±0.010 mg/L ±7% of reading at 25 °CTypical EMC ±0.001

Strona 39 - Molybdenum

44Nickel LR• Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of HI 9

Strona 40

45INTERFERENCESInterference may be caused by:Co2+ must not be presentFe2+ must not be presentAl3+ above 32 mg/LCa2+ above 1000 mg/L (as CaCO3)Cd2+ abo

Strona 41 - NICKEL HIGH RANGE

46NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source

Strona 42

47Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 50 s

Strona 43 - NICKEL LOW RANGE

48Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of reading at 25 °CTypical EMC ± 0.1 mg/LDeviationLigh

Strona 44 - Nickel LR

49Dissolved Oxygen• Let the sample stand and the flocculent agent will start to settle.• After approximately 2 minutes, when the upper half of the bot

Strona 45

5Where:-log I/Io= Absorbance (A)Io= intensity of incident light beamI = intensity of light beam after absorptionελ= molar extinction coefficient at wa

Strona 46 - MEASUREMENT PROCEDURE

50pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow ba

Strona 47

51pH• Reinsert the cuvette into the instrument.• Press the Read key to start the reading. The instrument displays the pH value.• Remove the cuvette an

Strona 48 - DISSOLVED OXYGEN

52Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.1

Strona 49

53Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until completely di

Strona 50

54Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC Dev.

Strona 51

55Phosphate LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternativ

Strona 52 - PHOSPHATE HIGH RANGE

56PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.2 mg/LLight

Strona 53

57Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the cap andshake gently until comp

Strona 54 - PHOSPHATE LOW RANGE

58SilverSILVERSPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.005 mg/LAccuracy ±0.020 mg/L ±5% of reading at 25 °CTypical EMC ± 0.001 mg/LDeviatio

Strona 55

59Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the secondbeaker (the sample) and swirl gently to mix. Press Timer and thedisplay will s

Strona 56 - PHOSPHORUS

6INSTRUMENT DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD).5) Splash proof keypad.6) ON/OFF power switc

Strona 57

60Silver• Fill a second cuvette up to the mark with 10 mL of the reacted sample.• Add 3 drops of Dispersing Agent (HI 93703-51),replace the cap and in

Strona 58

61ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Sou

Strona 59

62INTERFERENCESInterference may be caused by:Aluminum above 6 mg/LCadmium above 0.5 mg/LCopper above 5 mg/LIron above 7 mg/LManganese above 5 mg/LNick

Strona 60

63The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messages

Strona 61 - REQUIRED REAGENT

64The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.DATA MANAGEMENTDescription Range MethodAluminum 0.00 to

Strona 62

65ACCESSORIESREAGENT SETSHI 93701-01 100 free chlorine tests (powder)HI 93701-03 300 free chlorine tests (powder)HI 93701-F 300 free chlorine tests (l

Strona 63 - ERRORS AND WARNINGS

66OTHER ACCESSORIESHI 731318 cloth for wiping cuvettes (4 pcs)HI 731321 glass cuvettes (4 pcs)HI 731325W new cap for cuvette (4 pcs)HI 740034 cap for

Strona 64 - STANDARD METHODS

67WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and

Strona 65 - ACCESSORIES

68Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Strona 66

7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above

Strona 67 - HANNA LITERATURE

8The instructions listed below should be carefully followed during testing to ensure most accurate results.• Color or suspended matter in large amount

Strona 68 - MAN83207 04/10

9• Proper use of the powder reagent packet:(a) use scissors to open the powder packet;(b) push the edges of the packet to form a spout;(c) pour out th

Powiązane modele: HI 83207

Komentarze do niniejszej Instrukcji

Brak uwag