58
Silver
SILVER
SPECIFICATIONS
Range 0.000 to 1.000 mg/L
Resolution 0.005 mg/L
Accuracy ±0.020 mg/L ±5% of reading at 25 °C
Typical EMC ± 0.001 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm.
Method Adaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93737A-0 Buffer Reagent A 1 mL
HI 93737B-0 Buffer Reagent B 1 mL
HI 93737C-0 Indicator Reagent C 2 mL
HI 93737D-0 Fixing Reagent D 2 mL
HI 93703-51 Dispersing Agent 4-6 drops
(only when necessary, see note)
REAGENT SETS
HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
Select the
Silver
method using the procedure described in the
Method Selection
section (see page 12).
Note: for best results perform your tests between 20-24°C.
Fill two graduated beakers with 25 mL of sample.
Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker (the
blank) and swirl gently to mix.
# 1 # 2
# 1