pH
pH
SPECIFICATIONS
Range 6.5 to 8.5 pH
Resolution 0.1 pH
Accuracy ±0.1 pH
Typical EMC ±0.1 pH
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93710-0 Phenol Red Indicator 5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
pH
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter
is zeroed and ready for measurement.
Remove the cuvette and add 5 drops of HI 93710-0 Phenol
Red Indicator. Replace the cap and mix the solution.
10 mL
x 5
Dissolved Oxygen
Let the sample stand and the flocculent agent will start to settle.
After approximately 2 minutes, when the upper half of the bottle becomes
limpid, add 10 drops of HI 93732C-0.
Replace the cap and invert the bottle until the settled flocculent dissolves completely. The sample is
ready for measurement when it is yellow and
completely limpid.
Fill the cuvette up to the mark with 10 mL of the unreacted (original)
sample, and replace the cap. This is the blank.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when
the meter is zeroed and ready for measurement.
Remove the cuvette.
Fill another cuvette up to the mark with 10 mL of the reacted sample and
replace the cap.
Reinsert the cuvette into the instrument.
Press READ to start the reading. The instrument will display the results in mg/L of dissolved oxygen.
10 mL
x 10
INTERFERENCES
Interferences may be caused by reducing and oxidizing materials.