Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the vial.
Remove the cap and add 4 drops of HI 93764-0 Nessler
Reagent.
Replace the cap tightly and mix by inverting the vial a couple
of times. This is the sample.
Place the vial into the holder and push it completely down.
Press Timer and the display will show the countdown prior to the
measurement and the message “Reaction Time”. Alternatively, wait
for 3 minutes and 30 seconds and press Read. The instrument will
perform the reading.
The instrument displays concentration in mg/L of ammonia nitrogen (NH
3
-N). Press s or t to
convert the result in mg/L ammonia (NH
3
).
INTERFERENCES
Interference may be caused by:
Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine or urea above 10 ppm.
To eliminate these interferences distillation is required.
Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone above 0.1 %.
To eliminate these interferences distillation is required.
Sulfide: may cause turbidity.
Ammonia LR
4