5
Range 0.00-1.50 mg/L
Resolution 0.05 mg/L
Precision SD ±0.05 mg/L @ 0.50 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 560 nm
Light Detector Silicon Photocell
Method The reaction between Copper and the reagents causes a purple tint in
the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 4 x 1,5 volt AA batteries / 12 to 20 VDC through voltage adapter
Auto-Shut off
After 15' of non-use in measurement mode.
Dimensions 225 x 85 x 80 mm (8.7 x 3.3 x 3.1")
Weight 500 g (17,6 oz.).
REQUIRED REAGENTS
Code Description Quantity/test
HI 83740A-0 Copper Reagent A 5 mL
HI 83740B-0 Copper Reagent B 5 mL
HI 83740C-0 Copper Reagent C 2 x 4 spoons
HI 83740D-0 Copper Reagent D 2 x 4 spoons
SPECIFICATIONS
Precision is how closely repeated measurements agree
with each other. Precision is usually expressed as
standard deviation (SD). Accuracy is defined as the
nearness of a test result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
In a laboratory using a standard solution of 0.50
mg/L copper and a representative lot of reagent, an
operator obtained with a single instrument a standard
deviation of 0.05 mg/L.
PRECISION AND ACCURACY
16
Insert the zero (cuvet #1) into the holder and
close the lit.
#1
(Zero)
#2
(Sample)
Press ZERO and “----” will blink on the display.
After a few seconds the display will show “-0.0-”. The
meter is now zeroed and ready for measurement
Note:
If the “L Lo” (Low Light) message appears, the sample
must be diluted. See “General tips for an accurate
measurement” (page 12).
Remove the cuvet from the instrument.
Insert the reacted sample (cuvet #2) into the holder
and close the lid.
Press READ and the display will show “----” during
measurement.
The instrument directly displays concentration in mg/L
(ppm) of copper on the Liquid Crystal Display.
Note:
If the copper concentration exceeds 1.50 ppm or if
the sample is very turbid or dark red colored, it is
recommended to dilute the sample 10 times with
HI 83742-1 Wine Solvent-1 and repeat the
complete measurement procedure starting from the
beginning, taking 15 mL of diluted wine sample in
vial #1 for zero-sampling and 15 mL in vial #2
for reading. In this case the displayed value needs
to be multiplied by 10 to compensate for dilution.