This Instrument is in
Compliance with CE Directives
Instruction Manual
HI 95729C
Fluoride Low Range ISM
MAN95729CR2 10/04
www.hannainst.com
Przeglądanie stron 0
1 2 3 4 5 6 7 8 9 10

Podsumowanie treści

Strona 1 - Fluoride Low Range ISM

www.hannainst.comThis Instrument is inCompliance with CE DirectivesInstruction ManualHI 95729CFluoride Low Range ISMMAN95729CR2 10/04www.hannainst.c

Strona 2 - WARRANTY

10ERROR MESSAGESa) on zero readingThis indicates that the zeroing procedure failed due to a low signal-to-noise ratio. In this case press ZERO again.

Strona 3 - ACCESSORIES

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Strona 4 - GENERAL DESCRIPTION

3Please examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, please notify your Dealer

Strona 5 - CALIBRATION PROCEDURE

4The HI 95729 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru

Strona 6 - VALIDATION PROCEDURE

5Range 0.00 to 2.00 mg/LResolution 0.01 mg/LPrecision±0.03 mg/L @ 1.00 mg/LTypical EMC Deviation ±0.01 mg/LLight Source Tungsten lamp with narrow band

Strona 7 - FUNCTIONAL DESCRIPTION

6Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some o

Strona 8 - GUIDE TO DISPLAY CODES

7A microprocessor controlled special tungsten lamp emits radiation which is first opticallyconditioned and beamed to the sample contained in the cuvet

Strona 9 - MEASUREMENT PROCEDURE

8KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: to zero the meter prior to measurement.• READ/TIMER: this is a bi-functional key, ju

Strona 10

9This indicates that the meter is in a zeroed state and measurementcan be performed.Light over range: the cuvet is not inserted correctly and an exces