When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
Note: Pay attention that the selected range
is the one desired to run measurements.
If not, change the range following the
instructions from Range Selection.
pH MEASUREMENT
Fill the cuvette with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Remove the cuvette and add 5 drops of
the HI 93710-0 Phenol Red Indicator.
Replace the cap and shake gently the
solution.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press READ/ /TIMER. The lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.