Hanna-instruments HI 96104 Instrukcja Użytkownika Strona 21

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 28
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 20
21
CALIBRATION PROCEDURE
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or
ON/OFF keys.
Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
When calibrating, only the selected range is affected.
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Press and hold CAL CHECK for three
seconds to enter
calibration mode
. The
display will show “CAL” during
calibration procedure. The blinking
“ZERO” asks for instrument zeroing.
Place the CAL CHECK
TM
Standard
HI 96710-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for calibration. The blinking
“READ” asks for reading calibration
standard.
Remove the cuvette.
Przeglądanie stron 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 27 28

Komentarze do niniejszej Instrukcji

Brak uwag