Hanna-instruments HI 98170 Instrukcja Użytkownika Strona 36

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 35
36
LODIALL Requests all ISE Log on demand (HI 98172).
The answer string for each record contains:
The logged mode (2 chars):
01 - pH range (0.01 resolution)
02 - pH range (0.1 resolution)
03 - mV range
04 - Rel mV range
05 - ISE range
Reading status (1 char): R, O, U
Calculated reading, with sign and decimal point (7 chars) -
for pH, Rel mV and ISE range
Temperature reading, with sign and two decimal points
(7 chars)
mV reading status (1 char): R, O, U - only for HI 98171
and HI 98172
The mV reading, with sign and decimal point (7 chars) -
only for HI 98171 and HI 98172
The logged time, yymmddhhmmss (12 chars)
The calibration slope, with sign and decimal point (7 chars)
- not available for Rel mV range
The calibration offset, with sign and decimal point (7 chars) -
not available for ISE
Temperature probe presence (1 char)
Notes: “Err8” is sent if the instrument is not in measurement mode.
“Err6” is sent if the requested range is not available.
“Err4” is sent if the requested set parameter is not available.
“Err3” is sent if the Log on demand is empty.
“Err9” is sent if the battery power is less than 30%.
Invalid commands will be ignored.