Instruction Manual
HI 9126
Calibration Check
Waterproof
pH/mV/ºC Meter
www.hannainst.com
Przeglądanie stron 0
1 2 3 4 5 6 ... 16 17

Podsumowanie treści

Strona 1 - Waterproof

1Instruction ManualHI 9126Calibration CheckWaterproofpH/mV/ºC Meterwww.hannainst.com

Strona 2

18 19HI 9126 has beenprecalibrated for temperature at the factory.For optimum accuracy, it is recommended to recalibrate the meter fortemperature at l

Strona 3

20 21TAGS & SYMBOLS• Mode tags light up for indicating the corresponding active mode,and blink for warning the user.SETUP on: setup menu mode has

Strona 4 - OPERATIONAL GUIDE

22 23EEEEELECTRODE CONDITIONING & MAINTENANCELECTRODE CONDITIONING & MAINTENANCELECTRODE CONDITIONING & MAINTENANCELECTRODE CONDITIONING &

Strona 5

24 25TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnop

Strona 6

26 27pH ELECTRODESAll electrodes part numbers ending in B are supplied with a BNCconnector and 1 m (3.3') cable, as shown below:HI 1043BGlass-bod

Strona 7

28 29120 mm4.7"12 mm0.5"FC 10075 mm2.95"6 mm0.25"FC 200120 mm4.7"12 mm0.5"FC 210120 mm4.7"12 mm0.5"9.5mm DIA0.

Strona 8

30 31Hanna Instruments reserves the right to modify the design,construction and appearance of its products without advance notice.RECOMMENDATIONS FOR

Strona 9 - SETUP MENU

32Printed in EUROPE(ROMANIA)MAN9126Y 09/10Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Strona 10

2 3GENERAL DESCRIPTIONHI 9126 is a state-of-the-art waterproof, heavy-duty pH meter designedto provide laboratory results and accuracy under harsh ind

Strona 11 - LCD MESSAGE GUIDE

4 51) Liquid Crystal Display (LCD).2) RANGE key, to select pH or mV.3) ON/OFF key, to turn the meter ON and OFF.4) CAL key, to enter or exit calibrati

Strona 12

6 7INITIAL PREPARATIONThe meter is supplied with batteries (for details see page 19).To prepare the instrument for use, connect the pH electrode and t

Strona 13 - TROUBLESHOOTING GUIDE

8 9MEM & MR FUNCTIONSPressing the MEM key will memorize the reading on the display andstore it in the internal memory. The pH (mV) and temperature

Strona 14 - ACCESSORIES

10 11• Press the CFM key to confirm the calibration: the meter stores thefirst calibration point; the primary LCD will show the calibratedreading, whi

Strona 15 - 0.25"

12 13ONE-POINT CALIBRATIONFor optimum accuracy it is always recommended to perform a two-point calibration, but for a faster operation it is also poss

Strona 16 - ORP ELECTRODES

14 15The temperature has an effect on pH. The calibration buffer solutionsare affected by temperature changes to a lesser degree than normalsolutions.

Strona 17 - MAN9126Y 09/10

16 17The instrument allows the user to configure several parametersthrough the Setup Menu.To enter the Setup Menu, while in measurement mode, press an