TAGS & SYMBOLS
Mode tags light up for indicating the corresponding active mode,
and blink for warning the user.
SETUP on: setup menu mode has been entered.
MEASURE on: measurement mode.
CALIBRATION on: calibration mode has been entered.
MEMORIZE on: measurement stored in the internal memory and
frozen on the display
RECALL MEMORIZED on: stored value recalled.
Indication of temperature compensation mode:
MTC for manual, ATC for automatic compensation.
Battery symbol blinking: low battery condition. Batteries should
be replaced.
Calibration messages.
WAIT NOT READY blinking: buffer has been recognized, but
reading is not stable.
READY on: buffer has been recognized and reading is stable.
WRONG and WRONG blinking alternatively: wrong
buffer, value not recognized.
CLEAN ELECTRODE blinking: an abnormal difference between
new and previous calibration has been detected. Electrode clean-
ing is suggested. Follow the cleaning procedure described in the
“Electrode conditioning & maintenance” section. If the problem
remains, check the buffer solutions.
LCD MESSAGE GUIDELCD MESSAGE GUIDE
LCD MESSAGE GUIDELCD MESSAGE GUIDE
LCD MESSAGE GUIDE
To replace the batteries, follow the next steps:
Turn OFF the instrument.
Open the battery compartment cap (located on the top of the
instrument).
Remove old batteries.
Insert three new 1.5V AAA batteries in the battery compartment,
observing the polarity on the rear of the instrument.
Reattach the battery compartment cap.