Ammonia MR
SPECIFICATIONS
Range 0.00 to 10.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.05 mg/L ±5% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93715A-0 First Reagent 4 drops (6 drops for seawater)
HI 93715B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Ammonia MR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter is
zeroed and ready for measurement.
AMMONIA MEDIUM RANGE
Remove the cuvette.
Add carefully the content of one packet of HI 93755-0 Alkalinity
Indicator Reagent. Replace the cap and shake vigorously for 30
seconds.
Note: Pay attention not to spill reagent otherwise full color
development may be inhibited.
Replace the cuvette into the holder and close the lid.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ. When the timer ends the meter will
perform the reading.
The instrument displays the results in mg/L of alkalinity (CaCO
3
).
Alkalinity