Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press READ.
When the timer ends the meter will perform the reading. The instrument displays concentration in
µg/L of chromium VI.
Press the or to access the second level of functions.
Press the Chem Frm functional key to convert the result in µg/L of Chromate (CrO
4
2-
) and Dichromate
(Cr
2
O
7
2-
).
Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed
Iron above 1 ppm
Mercurous and mercuric ions cause slight inhibition of the reaction.
Chromium VI HR
Chromium VI HR
CHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range 0 to 1000 µg/L
Resolution 1 µg/L
Accuracy ±5 µg/L ±4% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1687-92
,
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93723-0 Powder reagent 1 packet
REAGENT SETS
HI 93723-01 Reagents for 100 tests
HI 93723-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the
Chromium VI HR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette up to the mark with 10 mL of unreacted sample and
replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The meter will show “-0.0-” when the meter is
zeroed and ready for measurement.
Remove the cuvette and add the content of one packet of
HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.
10 mL