20
Remove the cuvette.
Add 4 drops of HI 93715A-0 First Reagent (6 drops for
seawater analysis). Replace the cap and mix the solution.
Add 4 drops of HI 93715B-0 Second Reagent (10 drops for
seawater analysis). Replace the cap and mix the solution.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of ammonia
nitrogen (NH
3
-N).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of ammonia (NH
3
) and ammonium (NH
4
+
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various
aliphatic and aromatic amines.