ADDING REAGENT (COLORIMETRY only)
Because the reagent quantity is set up to react with 10 mL of
sample, it is very important to fill the cuvette correctly. The
liquid in the cuvette forms a meniscus; the bottom of the meniscus
must be at the same level with as the 10 mL mark.
To open the powder reagent pack:
a) use scissors to open the powder packet
b) push the edges of the packet to form a spout
c) pour the contents of the packet
Do not let the reacted sample stand too long after reagent is added or accuracy will be lost.
All the reaction times reported in this manual were recorded at 20 ºC (68 ºF). As a general rule
of thumb, they should be doubled at 10 ºC (50 ºF) and halved at 30 ºC (86 ºF).
Insert the cuvette with the mark aligned with the mark on the instrument top.
It is possible to take multiple readings in a row but it is recommended to take a new zero reading
for each sample and to use the same cuvette for zeroing and measurement.
After the reading, it is important to discard the sample immediately, otherwise the glass might
become permanently stained.
Note: In order to maximize accuracy prior to a measurement, follow the validation procedure to
be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.