Hanna-instruments HI 83414 Instrukcja Użytkownika Strona 41

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 44
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 40
41
LAMP REPLACEMENT
The instrument tungsten lamp has a life longer than 100,000 measurements. In case of lamp
failure, the defective lamp can be easily replaced. When the lamp is broken, the instrument displays
a lamp error message.
To replace the lamp follow the next steps:
Power off the instrument and take out the
power cord.
Remove the fixing screw of the lamp lid.
Remove the lamp lid.
Unscrew the lamp leads from connector.
Unlock the lamp and extract it by pulling it out
from the lamp holder.
Place the new lamp in the right position and
push it until is securely locked.
Insert the lamp leads into the connector and
tighten them using a screwdriver.
Return the lamp lid and secure with the screw.
Warning: After lamp replacement the meter has
to be recalibrated.
FUSE REPLACEMENT
To change the fuse follow next steps:
Disconnect the power cord from the rear panel of the instrument.
Pull out the fuse holder located near the power cord connector.
Replace the fuse with a similar one.
Push the fuse holder with the fuse in the appropriate place.
Przeglądanie stron 40
1 2 ... 36 37 38 39 40 41 42 43 44

Komentarze do niniejszej Instrukcji

Brak uwag