5-10
METHODS
5.5.3.1 Linear Dosing
This type of dosing uses equal volume doses of titrant as shown in the graph below.
The Linear Dosing option is recommended for slower, difficult nonaqueous titrations or
certain specific analysis.
Note: For steep and normal titration curves, smaller volume increments are recommended, to
obtain many measured points around the equivalence point.
For flat titration curves, larger volume increments are recommended for equivalence
point detection.
To set the dosing volume, select Linear Dosing option and type the optimum dose.
Linear Dosing
Przeglądanie stron 49