METHODS
5.5.18 Maximum Titrant Volume
The maximum titrant volume used in the titration must be set according to the analysis.
If the titration end point (fixed or equivalence End Point) is not reached, the titration will be
abnormally terminated after the maximum titrant volume has been dispensed. The error
message (Limits Exceeded) will appear on the display.
Range is from 0.100 to 100.000 mL.
5.5.19 Stirring Speed
The stirring speed can be set between 100 and 2500 RPM with 100 RPM resolution.
The preset value of stirrer speed from the current method is used during the entire process,
as long as the method is set as active.
Also, the speed can be adjusted using the and keys when the stirrer is on.