Strona 2 - TABLE OF CONTENTS

10Under range: A blinking “0.000” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sa

Strona 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Strona 4 - ABBREVIATIONS

12• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor dirt. Wipe it thorough

Strona 5 - PRECISION AND ACCURACY

13To compensate the meter for the sample turbidity or color, the measurement takes place in twophases. First, the meter is zeroed using the unreacted

Strona 6 - PRINCIPLE OF OPERATION

14• Remove the cuvette.• Add the content of one packet ofHI 93707-0 Nitrite Low Range reagent.Replace the cap and shake gently for15 seconds.• Replace

Strona 7 - FUNCTIONAL DESCRIPTION

15INTERFERENCESHighly buffered samples (alkalinity above 500 ppm) or with extreme pH may exceed thebuffering capacity of the reagent: neutralize acidi

Strona 8

16The reading should be within specifications as reported in the CAL CHECKTM StandardCertificate. If the value is found out of the specifications, ple

Strona 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Strona 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed andready for calibration. The blinking“READ” asks for reading calibratio

Strona 11

19• Press GLP/ to edit the desired year(2009-2099). If the key is kept pressed,the year number is automaticallyincreased.• When the correct year has

Strona 12

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DE

Strona 13 - MEASUREMENT PROCEDURE

20In the GLP mode, the last user calibration date can be verified and the factory calibration can berestored.To display the calibration date:• Press G

Strona 14

21• Press READ/ /TIMER to enter in thefactory calibration restore screen. Theinstrument asks for confirmation of usercalibration delete.• Press ZERO/C

Strona 15 - VALIDATION PROCEDURE

22To replace the instrument’s battery, follow the steps:• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the

Strona 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Strona 17 - CALIBRATION PROCEDURE

24Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Strona 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Strona 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96707 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Strona 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.000 to 0.600 mg/L (as NO2¯-N)Resolution 0.001 mg/LAccuracy ±0.020 mg/L ±4% of reading @ 25°CTypical EMC D

Strona 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Strona 22 - ACCESORIES

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette ho

Strona 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me

Strona 24 - MAN96707 12/10

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs

Komentarze do niniejszej Instrukcji

Brak uwag