Press CAL. The instrument will display the measured pH, the LCD first
expected buffer and the temperature reading.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, CFM functional
key is displayed.
Press CFM to confirm first point.
The calibrated value and the second expected buffer value is then dis-
played on the LCD.
After the first calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.