Hanna-instruments HI 98180 Instrukcja Użytkownika Strona 56

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 55
56
The ion charge (2 chars)
The ISE unit (2 chars)
The short name of the selected language (3 chars)
NSLx Requests the number of logged samples (4 chars).
The command parameter (1 char):
P - request for pH range
M - request for mV and Rel mV ranges
I - request for ISE range
LODPxxx Requests the xxx
th
pH record logged data.
LODMxxx Requests the xxx
th
mV/Rel mV record logged data.
LODIxxx Requests the xxx
th
ISE record logged data (HI 98184, HI 98185).
LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.
LODIALL Requests all ISE Log on demand (HI 98184, HI 98185).
The answer string for each record contains:
The logged mode (2 chars):
00 - pH range (0.001 resolution)
01 - pH range (0.01 resolution)
02 - pH range (0.1 resolution)
03 - mV range
04 - Rel mV range
05 - ISE range
Reading status (1 char): R, O, U
Calculated reading, with sign and decimal point and expo-
nent (11 chars) - for pH, Rel mV and ISE range
Temperature reading, with sign and two decimal points
(7 chars)
mV reading status (1 char): R, O, U
The mV reading, with sign and decimal point (7 chars)
The logged time, yymmddhhmmss (12 chars)
The calibration slope, with sign and decimal point (7 chars)
- not available for Rel mV range
The calibration offset, with sign and decimal point (7 chars) -
not available for ISE
Temperature probe presence (1 char)