The calibrated value will be displayed on the primary LCD and the
second expected buffer value on the secondary LCD.
After the first calibration point is confirmed, submerse the pH
electrode and the temperature probe approximately 3 cm (1¼”) into
the second buffer solution and stir gently. The temperature probe
should be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected
buffer, the “READY” tag will be displayed and the
CFM” tag will blink.
Press CFM to confirm calibration.
The calibrated value is then displayed on the primary
LCD and the third expected buffer value on the
secondary LCD.
After the second calibration point is confirmed, submerse the pH
electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe
should be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is
stable.
When the reading is stable and close to the selected
buffer, the “READY” tag will be displayed and the
CFM” tag will blink.
Press CFM to confirm calibration.
The calibrated value is then displayed on the primary LCD and the
fourth expected value on the secondary LCD.
After the third calibration point is confirmed, submerse the pH electrode
and the temperature probe approximately 3 cm (1¼”) into the
next buffer solution and stir gently. The temperature probe should
be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.