Hanna-instruments HI 2214 Instrukcja Użytkownika Strona 14

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 13
14
Press the ARROW keys to select another buffer to be
replaced.
Press CFM to confirm the buffer that will be replaced.
Press CAL to leave calibration without replacing.
Note: If the replaced buffer is outside the ±0.2 pH
window, around each of the calibrated buffers, it is
possible to select this buffer for next calibration
during current calibration.
WORKING WITH CUSTOM BUFFERS
If a custom buffer was set in SETUP menu, it can be selected during
calibration by pressing the ARROW keys. The “BUFFER pH” tag will
blink.
Press SETUP if you want to adjust the buffer value. The buffer value will
start blinking.
Use the ARROW keys to change the buffer value.
After 5 seconds the buffer value is updated. Press SETUP if you want to
change it again.
Note: Custom buffer value can be adjusted in a ±1.00 pH window,
around the set value.
Przeglądanie stron 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 47 48

Komentarze do niniejszej Instrukcji

Brak uwag