Strona 1 - HI 83218

1Instruction ManualHI 83218Multiparameter BenchPhotometerfor Laboratorieswww.hannainst.com

Strona 2 - TABLE OF CONTENTS

10• Do not let the reacted sample stand too long after reagent is added. For best accuracy, respect thetimings described in each specific method.• It

Strona 3 - ABBREVIATIONS

11POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note: Alw

Strona 4 - PRINCIPLE OF OPERATION

12DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measurem

Strona 5

13In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that ch

Strona 6 - FUNCTIONAL DESCRIPTION

14Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the W X functional keys to h

Strona 7 - KEYPAD DESCRIPTION

15HI 83218 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will disp

Strona 8

16SPECIFICATIONSRange 0.0 to 50.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±5% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source Tungsten

Strona 9 - USING CUVETTES

17Ammonia HR• Remove the cuvette.• Add 4 drops of HI 93733A-0 Nessler Reagent. Replacethe cap and mix the solution.• Reinsert the cuvette into the ins

Strona 10 - HEALTH & SAFETY

1810 mLAmmonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationL

Strona 11 - OPERATIONAL GUIDE

19• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add 4 drops o

Strona 12 - SPECIAL CONVERSIONS

2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will prov

Strona 13 - Contrast

20Chromium VI HRCHROMIUM VI HIGH RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±5 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/LDe

Strona 14

21• Remove the cuvette and add the content of one packet ofHI 93723-0 reagent. Replace the cap and shake vigorouslyfor about 10 seconds.• Reinsert the

Strona 15 - HELP MODE

2210 mLChromium VI LRCHROMIUM VI LOW RANGESPECIFICATIONSRange 0 to 300 μg/LResolution 1 μg/LAccuracy ±1 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/

Strona 16 - AMMONIA HIGH RANGE

23• Remove the cuvette and add the content of one packet ofHI 93749-0 reagent. Replace the cap and shake vigorouslyfor about 10 seconds.• Reinsert the

Strona 17 - INTERFERENCES

24NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source

Strona 18 - AMMONIA MEDIUM RANGE

25Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 50 s

Strona 19

2610 mLNitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of reading at 25 °CTypical EMC ±1 mg/LDeviat

Strona 20 - CHROMIUM VI HIGH RANGE

27Nitrite HR• Add the content of one packet of HI 93708-0 reagent.Replace the cap and shake gently until completely dissolved.• Reinsert the cuvette i

Strona 21

2810 mLSPECIFICATIONSRange 0.00 to 1.15 mg/LResolution 0.01 mg/LAccuracy ±0.06 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Source

Strona 22 - CHROMIUM VI LOW RANGE

29• Add the content of one packet of HI 93707-0 reagent.Replace the cap and shake gently for about 15 seconds.• Reinsert the cuvette into the instrume

Strona 23

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Strona 24 - MEASUREMENT PROCEDURE

30PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.2 mg/LLight

Strona 25

31• Add the content of one packet of HI 93706B-0Phosphorus Reagent B (Amino Acid) to the cuvette.Replace the cap and shake gently until completelydiss

Strona 26 - NITRITE HIGH RANGE

32The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messages

Strona 27 - Nitrite HR

33Description Range MethodAmmonia HR 0.0 to 50.0 mg/L NesslerAmmonia MR 0.00 to 10.00 mg/L NesslerChromium VI HR 0 to 1000 μg/L Diphenylcarbohydrazide

Strona 28 - NITRITE LOW RANGE

34OTHER ACCESSORIESHI 731318 cloth for wiping cuvettes (4 pcs)HI 731321 glass cuvettes (4 pcs)HI 731325W new cap for cuvette (4 pcs)HI 740034 cap for

Strona 29

35WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and

Strona 30 - PHOSPHORUS

36Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Strona 31

4PRECISION AND ACCURACYPrecision is how closely repeated measurements agreewith each other. Precision is usually expressed asstandard deviation (SD).A

Strona 32 - ERRORS AND WARNINGS

5Where:-log I/Io= Absorbance (A)Io= intensity of incident light beamI = intensity of light beam after absorptionελ= molar extinction coefficient at wa

Strona 33 - STANDARD METHODS

6INSTRUMENT DESCRIPTIONFUNCTIONAL DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6

Strona 34 - ACCESSORIES

7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above

Strona 35 - HANNA LITERATURE

8(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hangingon the tip of the syringe, if so eliminate them.

Strona 36 - MAN83218 04/10

9USING CUVETTES• Proper mixing of the cuvette is done by shaking the cuvette, moving the cuvette up and down. Themovement may be gentle or vigorous. T