10 mL
SPECIFICATIONS
Range 0.00 to 1.15 mg/L
Resolution 0.01 mg/L
Accuracy ±0.06 mg/L ±4% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
EPA Diazotization method 354.1
. The reaction between nitrite and
the reagent causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93707-0 Powder reagent 1 packet
REAGENT SETS
HI 93707-01 Reagents for 100 tests
HI 93707-03 Reagents for 300 tests
For other accessories see page 34.
MEASUREMENT PROCEDURE
Select the
Nitrite LR
method using the procedure described in the
Method Selection
section (see
page 11).
Fill the cuvette up to the mark with 10 mL of unreacted sample
(up to the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Nitrite LR
NITRITE LOW RANGE