Add the content of one packet of HI 93707-0 reagent.
Replace the cap and shake gently for about 15 seconds.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 6 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in mg/L of nitrite.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of nitrogen-nitrite (NO
2
¯-N) and sodium nitrite
(NaNO
2
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by the following ions:
ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric, lead, metavanadate and chloroplatinate.
Strongly reducing and oxidizing reagents.
High levels of nitrate (above 100 mg/L) could yield falsely high readings due to a minute amount of
reduction to nitrite that could occur at these levels.
Nitrite LR