43
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet
of HI 93750B-0 reagent. Replace the cap and gently
mix for one minute by slowly turning the cuvette upside
down.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L (ppm) of potassium (K).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
Press or to return to the measurement screen.
For ULTRA HIGH RANGE samples: follow the procedure described at page 44.
Potassium HR