SPECIFICATIONS
Range 0 to 100 mg/L
Resolution 5 mg/L
Accuracy ±5 mg/L ±3% of reading at 25 °C
Light Source Tungsten lamp with narrow band interference filter @ 466 nm
Method Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension
is measured.
REQUIRED REAGENTS
Code Description Quantity
HI 93751-0 Indicator reagent 1 packet
REAGENT SETS
HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 57.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure on page 20.
Select the
Sulfate
method using the procedure described
in the
Method Selection
section (see page 12).
Fill a cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0- when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Sulfate
SULFATE
10 mL